Need to know if there are any humidity and temperature requirements/specificications that come with the use of this equipment?

Hi,I have problem with GE Innova 2100Iiq. The system dispays collision on the detector. I have checked the collision circuitry - there is no collision and all collision switches operate normally. I suspect PID bd, but I'm not very sure. Has anyone had a similar problem?Thanks for your response.

Yêu cầu hoàn tiền nếu đơn hàng của bạn không được vận chuyển, bị thiếu hoặc giao đến có vấn đề về sản phẩm, cùng với trả hàng miễn phí về kho địa phương nếu bị lỗi

Could someone help me with: Unintended motion risk. Motion command failure. Move carefully (1.6) I can't find my error message list

Image

Attempting to repair GE collimator , part# 5127609 "Performed testing on Innova 2100 collimator. Collimator was found to be non-functional with numerous collimator power failure errors. Attempts were made to test cabling and run diagnostics, however collimator would not boot far enough to do so." Error Reads: 600524 - Under Table collimator error height limit switch 3400017- System boot successful 600206- Under Table collimator is offline 3400005-The System is booting 3400012- The system is resetting 600524-Under Table collimator error height limit switch 3400017-System boot successful 600206-Under Table collimator is offline

I'm working on an engineering study at my facility and am looking for the physical specifications for our GE Innova 2100 IQ .  Specifically, I'm looking for the weight, electrical requirements, and heating load.  Does anyone have a generic cut sheet for this machine?  Thank you.

Need to know if there are any humidity and temperature requirements/specificications that come with the use of this equipment?

Hi,I have problem with GE Innova 2100Iiq. The system dispays collision on the detector. I have checked the collision circuitry - there is no collision and all collision switches operate normally. I suspect PID bd, but I'm not very sure. Has anyone had a similar problem?Thanks for your response.

"Performed testing on Innova 2100 collimator. Collimator was found to be non-functional with numerous collimator power failure errors. Attempts were made to test cabling and run diagnostics, however collimator would not boot far enough to do so."

I'm working on an engineering study at my facility and am looking for the physical specifications for our GE Innova 2100 IQ .  Specifically, I'm looking for the weight, electrical requirements, and heating load.  Does anyone have a generic cut sheet for this machine?  Thank you.