Image

Hi everyone, I am writing you regarding the driver module fault on my GE Signa 1.5T HDXT system. It gives error like this.... DD Driver Fault Detected - Hihg Voltage Fault Source of Fault is BD4 Value =468 V Limit:475V Initially it was happening very rare - perhaps once in month during the last year, then every day and now a few time in a day. My first reaction was doing TPS reset. It was helping, but not all the times. Then I have swapped all cables between BD4 and DD path - starting from equipment room to PP and from PP to hybrid module. The result is same. It looks like to problem is in driver module -5110612. But I need to make sure it does. Could be problem related to software? Any clues? Thank you in advance.