Material & methods: A dedicated phantom (Mercury 4.0) was scanned using two of the most frequently used clinical scanning protocols (chest and abdomen-pelvis). The preset protocol settings were used as starting points (reference conditions). Scan direction, scan mode (helical vs. axial), total beam width, tube potential (kVp), and ATCM settings were then modified individually to understand their impact on radiation dose and image quality. Regarding the ATCM settings, the SmartmA minimum and maximum mA limits, and the noise index (NI) values were varied. As surrogates of patient dose, the CTDIvol and DLP values of each scan were used. As surrogates of image quality were used the image noise and the detectability index (d') of five different materials (air, solid water, polystyrene, iodine, and bone) embedded in the Mercury phantom calculated with the ImQuest software.
Results: The scanning direction did not have any effect on ATCM curves, unlike what has been observed in CT scanners from other manufacturers. Total beam width does matter, however, the SmartmA limit settings and kVp selection had the greatest impact on image quality and dose. It was seen that improper minimum mA limit settings practically invalidated the ATCM operation. In contrast, when full modulation was allowed without restrictions, noise standard deviation, and detectability index became much more consistent across the wide range of phantom diameters. For lower kVp settings an impressive dose reduction was observed that requires further investigation.
Solicite um reembolso se o seu pedido não for enviado, for extraviado ou chegar com problemas no produto, além de devoluções locais gratuitas para defeitos
Conclusion: SmartmA is a tool that if not properly used may increase the patient doses considerably. Therefore, its settings should be carefully adjusted for each preset different clinical protocol.
© 2024 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Purpose: To investigate the operation principles of the automatic tube current modulation (ATCM) of a modern GE healthcare CT scanner, and the impact of related settings on image quality and patient dose.