The principal perioperative issue in patients with implantable cardioverter-defibrillators (ICDs) is potential electromagnetic interference (EMI) generated from monopolar electrosurgery. This can result in inappropriate tachyarrhythmia detection and resultant therapy including shocks. To avoid this problem, anesthesia providers must be prepared to inactivate tachyarrhythmia therapies when EMI is likely. Suspension of tachyarrhythmia therapies can be done via manual device reprogramming or by placing a magnet on the device. Many providers use a magnet to suspend antitachycardia therapy; however, there are reports of failed magnet response and inappropriate shock with EMI.1–5 In addition, when using a magnet, many providers routinely position the magnet centrally over the device; however, some manufacturers recommend off-center magnet placement.
Contribution: This author helped interpret the data, draft the manuscript, approve the final version to be published, and ensure the accuracy and integrity of the work.
The Heart Rhythm Society (HRS) and ASA in conjunction with the AHA, ACC, and the Society of Thoracic Surgeons have developed a consensus statement regarding perioperative cardiac implantable electronic device (CIED) management, and in 2020, the ASA released an updated practice advisory.7,8 Patients with ICD should have documented device interrogation within 3 to 6 months of a surgical procedure to identify battery life, programmed therapies, pacing dependence, and tachyarrhythmias.7,8 There have been case reports of inappropriate ICD therapy, CIED damage, and patient injury with inadequate preoperative evaluation and erroneous assumptions about the effects of magnet application.1,2
Contribution: This author helped with conception, clinical data interpretation, and drafting and editing the manuscript; has approved the final version to be published; and ensured the accuracy and integrity of the work.
Costco trolley baby seat
In conclusion, we report a rare case of “appropriate” ICD detection and therapy for VF despite magnet placement. This case highlights the importance of accurate magnet placement using the recommended positioning from each device manufacturer. It is important for the anesthesiologist to ensure proper CIED function and obtain CIED-related recommendations before surgery.
When EMI is likely, magnet placement to suspend antitachycardia therapy may suffice in some patients with ICD that are not pacing-dependent. This presumption is not appropriate for all patients and care should be individualized.7–9 The most recent ASA practice advisory cautions against indiscriminate use of magnet over an ICD.8 While magnets prove useful in their ability to temporarily suspend tachyarrhythmia therapy, each device manufacturer has different specifications for proper magnet placement (Figure 2 and Table). For instance, a centrally positioned donut magnet may not maintain an adequate magnetic field to suspend tachyarrhythmia therapies on Abbott CIED. Instead, Abbott recommends placing the magnet to the side of the device in its long axis so that the arc of the magnet is over most of the generator (Figure 2A). Inaccurate magnet placement can result in inappropriate shock or life-threatening asystole during surgery. We suggest displaying a chart with appropriate magnet positioning in operating rooms, emergency rooms, or crash carts to avoid this situation.
Olympia folding cartprice
This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.
Gajendran, Vishal K. MD*; Contractor, Tahmeed MBBS†; Tone, Ryan C. MD*; Mascetti, Carin R. MD*; McCabe, Melissa D. MD, MSCR*
Reinforced steel plate composite wobble-free stabilizer steel. Toe plate folds down to fit into smallest possible places at home, car, workplace. Wobble-free composite support extends high to 40 in. Features smooth running rust-free wheels.
There are several other issues with reliance on magnet application for suppression of tachycardia therapies. Only Boston Scientific and Medtronic devices emit a tone when a magnet is detected (Table). However, continuous suspension of antitachyarrhythmia therapy can only be confirmed with Boston Scientific ICD that emits a beeping tone once every second for the duration of magnet detection. Auscultation of the generator with a stethoscope may enhance the provider’s ability to hear the tones when there is significant ambient noise or when the device is implanted in a deep pocket. The response to magnet application may be programmable, and in rare situations, magnet application will not suspend tachyarrhythmia therapies. For ICD with a programmable magnet response, the response to magnet application must be programmed “ON” for a magnet to suspend tachyarrhythmia therapy.
The patient was extubated and transported to the postanesthesia care unit. The ICD was interrogated, confirming ventricular arrhythmia and appropriate therapy (Figure 1A). Notably, magnet detection by the device was intermittent (Figure 1B) despite the magnet being secured in position without disruption. A detailed discussion with the cardiac electrophysiology team and Abbott technical services revealed that the cause of this phenomenon was inaccurate placement of the magnet centrally over the device.
Olympia folding cartreviews
Address correspondence to Melissa D. McCabe, MD, Department of Anesthesiology, Loma Linda University, 11234 Anderson St, MC 2532, Loma Linda, CA. Address e-mail to [email protected].
We report a case where, despite placement of a magnet over an ICD, it failed to suspend the tachyarrhythmia therapy and the patient received an appropriate shock for ventricular fibrillation (VF) shortly after induction of anesthesia. We discuss the reasons a response may not be elicited by a magnet and the magnet usage guidelines from each manufacturer. Health Insurance Portability and Accountability Act written authorization for publication of the case report was obtained from the patient.
A 52-year-old man with a prior history of cecal volvulus, right hemicolectomy, and loop ileostomy presented for ileostomy reversal. His medical history included coronary artery disease with a prior myocardial infarction, American College of Cardiology (ACC) class C heart failure with reduced ejection fraction (left ventricular ejection fraction of 25%), severe mitral regurgitation, and associated World Health Organization class 2 moderate pulmonary hypertension. An ICD (Fortify Assura VR 1357-40Q Abbott, formerly St Jude Medical, Sylmar, CA) with optisure bipolar lead was placed 8 months prior. It was programmed for the detection and therapy of VF at 200 bpm and to only monitor for ventricular tachycardia at 171 bpm. Bradycardia parameters were programmed to VVI 40 bpm. The patient completed an outpatient ICD interrogation 3 months prior. The underlying rhythm was sinus at 98 bpm, and there was no evidence of tachyarrhythmias or pacing dependence. He was on ACC/American Heart Association (AHA) guideline–directed medical therapy for heart failure and was felt to be euvolemic before surgery.6
Costco trolleyCart
The fundamental perioperative concern for patients with implantable cardioverter-defibrillators (ICDs) is the potential for electromagnetic interference (EMI) from monopolar electrosurgery. The ICD may interpret electromagnetic signals as a tachyarrhythmia and deliver an inappropriate shock to the patient. Magnet placement is often used to avoid this problem since a magnet will often deactivate an ICD’s tachyarrhythmia therapy. We report a case in which magnet placement over an ICD failed to suspend tachyarrhythmia therapy because of imprecise magnet positioning. This case demonstrates the possibility for error when relying on a magnet to suspend tachyarrhythmia therapies.
Before induction, standard American Society of Anesthesiologists (ASA) monitors were placed, and the radial artery was cannulated. Induction proceeded with lidocaine, propofol, fentanyl, and rocuronium, followed by endotracheal intubation. Phenylephrine was bolused as needed to maintain systolic blood pressure >130 mm Hg and mean arterial pressure 80 to 90 mm Hg. A magnet was placed centrally over the ICD generator and secured in place with tape. The patient was positioned supine, prepped, and draped with an electrosurgery dispersive electrode (COVIDIEN, Mansfield, MA) placed on the right outer thigh. Before incision, the patient developed VF and hemodynamic instability with loss of pulsatility in the arterial line waveform. Although the magnet remained secured over the generator, the ICD delivered appropriate therapy before the anesthesiologists attempted to remove the magnet. Hemodynamics were supported with 10-µg epinephrine and pressures normalized after normal sinus rhythm was restored. Anesthesiology and surgery teams decided to postpone the case pending cardiac electrophysiology evaluation.
MacSports XL SteelFoldingWagon
Magnets are sometimes placed on ICD at the time of surgery to avoid EMI-related inappropriate shocks. We report a case of a magnet failing to suppress ICD firing as a result of inaccurate positioning, serendipitously leading to an appropriate shock for VF that developed after induction of anesthesia in a high-risk patient. Perioperative pacemaker and ICD management by the anesthesia provider is crucial to avoid asystole in pacing-dependent patients or inappropriate shocks in patients with ICD due to electrosurgery-related EMI.
Contribution: This author helped draft and edit the manuscript, approve the final version to be published, and ensure the accuracy and integrity of the work.
Contribution: This author helped care the patient, draft and edit the manuscript, approve the final version to be published, and ensure the accuracy and integrity of the work.
Costco trolley uk
Contribution: This author helped care the patient, interpret the clinical data, draft and edit the manuscript, approve the final version to be published, and ensure the accuracy and integrity of the work.
While the primary perioperative concern with CIED remains EMI, in this case, the ICD appropriately recognized an intrinsic, aberrant rhythm, and delivered therapy. This case demonstrates the potential for errors with reliance on magnet use. One solution for patients undergoing general anesthesia is to have expert interrogation and programming of any CIED. However, this infrastructure is not well established in many institutions. While this patient did not experience harm, this will clearly not always be the case when EMI occurs. Therefore, familiarity with manufacturer recommendations and device functionality is prudent for today’s anesthesiologists.
GLOSSARY ACC American College of Cardiology AHA American Heart Association ASA American Society of Anesthesiologists BOL beginning of life CIED cardiac implantable electronic device; EMI electromagnetic interference ERI elective replacement indicator HRS Heart Rhythm Society ICD implantable cardioverter-defibrillator; VF ventricular fibrillation; VS ventricle sensed
Surgical positioning also affects magnet usage. Magnets are more likely to become dislodged when surgery is performed in the prone or lateral position. Inappropriate ICD shocks have been described when a magnet was used to suspend tachyarrhythmia therapy in the lateral position.2 In addition, patient habitus may complicate magnet placement. Inappropriate shock with EMI has been described in patients with high body mass index as it may be difficult to maintain an adequate magnetic field when there is excess tissue overlying the generator.4 Rarely, software incompatibility between the device and programmer may contribute to problems with reprogramming.3